ACCURACY OF T(d,n)4He NEUTRON ENERGY MEASUREMENTS USING AN NE-213 LIQUID SCINTILLATOR.

Alex Tsechanski, Gad Shani

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

A well-collimated T(d,n)**4He fact neutron beam has been used to investigate the possibility of a precise measurement of the energy of fast neutrons using a 2- multiplied by 2-in. NE-213 liquid scintillator. Four sets of measurements were performed at 0-, 85-, 90-, and 95-deg nominal angles of the deuteron beam relative to the collimator axis. This experimental setup provides monoenergetic neutrons with nominal energies of 14. 697, 14. 115, 14. 061, and 14. 007 Mev, respectively. The results of the energy measurement of these monoenergetic neutrons are presented in this paper. The proton recoil spectra created in the liquid scintillator were unfolded with the FORIST unfolding code.

Original languageEnglish
Pages (from-to)78-87
Number of pages10
JournalNuclear Technology
Volume64
Issue number1
DOIs
StatePublished - 1 Jan 1984

ASJC Scopus subject areas

  • Nuclear and High Energy Physics
  • Nuclear Energy and Engineering
  • Condensed Matter Physics

Fingerprint

Dive into the research topics of 'ACCURACY OF T(d,n)4He NEUTRON ENERGY MEASUREMENTS USING AN NE-213 LIQUID SCINTILLATOR.'. Together they form a unique fingerprint.