T1 - Beta ray spectroscopy based on a plastic scintillation detector/silicon surface barrier detector coincidence telescope
AU - Horowitz, Y. S.
AU - Hirning, C. R.
AU - Yuen, P.
AU - Aikens, M.
N1 - Funding Information:
This research was supported by the CANDU Owner's Group (COG) Health and Safety Program. We would also like to thank AECL Research Co . for the use of their facilities .
PY - 1994/1/15
Y1 - 1994/1/15
N2 - Beta radiation is now recognized as a significant radiation safety problem and several international conferences have recently been devoted to the problems of mixed field beta/photon dosimetry. Conventional dosimetry applies algorithms to thermoluminescence dosimetry (TLD) multi-element badges which attempt to extract dose information based on the comparison of TL signals from "thick/thin" and/or "bare/filtered" elements. These may be grossly innacurate due to inadequate or non-existant knowledge of the energy spectrum of both the beta radiation and the accompanying photon field, as well as other factors. In this paper, we discuss the operation of a beta-ray energy spectrometer based on a two element, E × dE detector telescope intended to support dose algorithms with beta spectral information. Beta energies are measured via a 5 cm diameter × 2 cm thick BC-404 plastic scintillator preceded by a single, 100 μm thick, totally depleted, silicon dE detector. Photon events in the E detector are rejected by requiring a coincidence between the E and dE detectors. Photon rejection ratios vary from 225:1 at 1.25 MeV (60Co) to 360:1 at 0.36 MeV (133Ba). The spectrometer is capable of measuring electron energies from a lower energy coincidence threshold of approximately 125 keV to an upper limit of 3.5 MeV. This energy range spans the great majority of beta-emitting radionuclides in nuclear facilities.
AB - Beta radiation is now recognized as a significant radiation safety problem and several international conferences have recently been devoted to the problems of mixed field beta/photon dosimetry. Conventional dosimetry applies algorithms to thermoluminescence dosimetry (TLD) multi-element badges which attempt to extract dose information based on the comparison of TL signals from "thick/thin" and/or "bare/filtered" elements. These may be grossly innacurate due to inadequate or non-existant knowledge of the energy spectrum of both the beta radiation and the accompanying photon field, as well as other factors. In this paper, we discuss the operation of a beta-ray energy spectrometer based on a two element, E × dE detector telescope intended to support dose algorithms with beta spectral information. Beta energies are measured via a 5 cm diameter × 2 cm thick BC-404 plastic scintillator preceded by a single, 100 μm thick, totally depleted, silicon dE detector. Photon events in the E detector are rejected by requiring a coincidence between the E and dE detectors. Photon rejection ratios vary from 225:1 at 1.25 MeV (60Co) to 360:1 at 0.36 MeV (133Ba). The spectrometer is capable of measuring electron energies from a lower energy coincidence threshold of approximately 125 keV to an upper limit of 3.5 MeV. This energy range spans the great majority of beta-emitting radionuclides in nuclear facilities.
UR - http://www.scopus.com/inward/record.url?scp=0028255763&partnerID=8YFLogxK
U2 - 10.1016/0168-9002(94)91337-4
DO - 10.1016/0168-9002(94)91337-4
M3 - Article
AN - SCOPUS:0028255763
SN - 0168-9002
VL - 338
SP - 522
EP - 533
JO - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
JF - Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
IS - 2-3
ER -