TY - JOUR
T1 - Chemiluminescent optical fiber immunosensor for detection of autoantibodies to ovarian and breast cancer-associated antigens
AU - Salama, Orly
AU - Herrmann, Sebastien
AU - Tziknovsky, Alina
AU - Piura, Benjamin
AU - Meirovich, Michael
AU - Trakht, Ilya
AU - Reed, Brent
AU - Lobel, Leslie I.
AU - Marks, Robert S.
PY - 2007/2/15
Y1 - 2007/2/15
N2 - We report herein the development of an optical fiber based chemiluminescent immunosensor for detection of the native autoimmune response to GIPC-1, a PDZ containing protein involved in regulation of G-protein signaling. The recombinant protein GIPC-1 was expressed in bacteria, purified, refolded and conjugated to the tip of an optical fiber. A human monoclonal 27.B1 IgM isolated from a breast cancer patient, which targets the GIPC-1 protein, was used for calibration of the immunosensor and was detected down to a concentration of 30 pg/ml. We determined that the fiber-optic immunosensor had a detection limit 50 times lower than chemiluminescent ELISA, and approximately 500 times lower than colorimetric ELISA. In addition, sera from 11 ovarian cancer patients, 22 breast cancer patients and asymptomatic controls were tested for the presence of IgM anti-GIPC-1 autoantibodies in their serum using the two methods. The immunosensor assay detected 54% and 77% GIPC-1 positive sera within ovarian and breast cancer patients, respectively, as compared to chemiluminescent ELISA, which only detected 18% and 27%, respectively. We envision that this immunosensor may serve as a diagnostic tool for screening women for ovarian and breast cancer at an early stage, thus increasing their chance of survival.
AB - We report herein the development of an optical fiber based chemiluminescent immunosensor for detection of the native autoimmune response to GIPC-1, a PDZ containing protein involved in regulation of G-protein signaling. The recombinant protein GIPC-1 was expressed in bacteria, purified, refolded and conjugated to the tip of an optical fiber. A human monoclonal 27.B1 IgM isolated from a breast cancer patient, which targets the GIPC-1 protein, was used for calibration of the immunosensor and was detected down to a concentration of 30 pg/ml. We determined that the fiber-optic immunosensor had a detection limit 50 times lower than chemiluminescent ELISA, and approximately 500 times lower than colorimetric ELISA. In addition, sera from 11 ovarian cancer patients, 22 breast cancer patients and asymptomatic controls were tested for the presence of IgM anti-GIPC-1 autoantibodies in their serum using the two methods. The immunosensor assay detected 54% and 77% GIPC-1 positive sera within ovarian and breast cancer patients, respectively, as compared to chemiluminescent ELISA, which only detected 18% and 27%, respectively. We envision that this immunosensor may serve as a diagnostic tool for screening women for ovarian and breast cancer at an early stage, thus increasing their chance of survival.
KW - Cancer-associated antigen
KW - Chemiluminescence
KW - GIPC-1
KW - Human monoclonal antibody
KW - Optical fiber immunosensor
KW - Ovarian and breast cancer
UR - http://www.scopus.com/inward/record.url?scp=33846386134&partnerID=8YFLogxK
U2 - 10.1016/j.bios.2006.07.003
DO - 10.1016/j.bios.2006.07.003
M3 - Article
C2 - 16904309
AN - SCOPUS:33846386134
SN - 0956-5663
VL - 22
SP - 1508
EP - 1516
JO - Biosensors and Bioelectronics
JF - Biosensors and Bioelectronics
IS - 7
ER -