Abstract
The Cherenkov counting efficiency varies strongly with color quenching, thus correction curves must be used to obtain correct results. The external 152Eu source of a Quantulus 1220TM liquid scintillation counting (LSC) system was used to obtain a quench indicative parameter based on spectra area ratio. A color quench correction curve for aqueous samples containing 90Sr/90Y was prepared. The main advantage of this method over the common spectra indicators is its usefulness also for low level Cherenkov counting.
Original language | English |
---|---|
Pages (from-to) | 805-808 |
Number of pages | 4 |
Journal | Applied Radiation and Isotopes |
Volume | 67 |
Issue number | 5 |
DOIs | |
State | Published - 1 May 2009 |
Keywords
- Cherenkov counting
- External source
- Quench correction