Abstract
A detailed kinetic analysis has been performed of the thermoluminescence (TL) glow curve of high purity synthetic quartz. The kinetic parameters of the glow peak at 110°C were evaluated for doses ranging from 0.1 Gy to 100 Gy using glow curve deconvolution (GCD), initial rise, variable heating rate and phosphorescence decay methods. All the methods gave results that agree within the experimental errors.
Original language | English |
---|---|
Pages (from-to) | 225-228 |
Number of pages | 4 |
Journal | Radiation Protection Dosimetry |
Volume | 100 |
Issue number | 1-4 |
DOIs | |
State | Published - 1 Jan 2002 |
Externally published | Yes |
ASJC Scopus subject areas
- Radiation
- Radiological and Ultrasound Technology
- Radiology Nuclear Medicine and imaging
- Public Health, Environmental and Occupational Health