Abstract
The effect of previous irradiation on the sensitivity of the glow peaks of LiF:Mg,Ti (TLD-100) is investigated up to levels of dose of 400 Gy in both slow-cooled and naturally cooled materials following the 400°C/1 hour pre-irradiation anneal. It is demonstrated that the naturally cooled samples can be re-used up to accumulated levels of dose of 50 Gy without recalibration. At 400 Gy a significant decrease in sensitivity of approximately 25% is observed for all the glow peaks (excluding peak 3). In slow-cooled materials even 100 Gy does not alter the sensitivity of the material.
Original language | English |
---|---|
Pages (from-to) | 232-237 |
Number of pages | 6 |
Journal | Radiation Protection Dosimetry |
Volume | 188 |
Issue number | 2 |
DOIs | |
State | Published - 13 Jun 2020 |
ASJC Scopus subject areas
- Radiological and Ultrasound Technology
- Radiation
- Radiology Nuclear Medicine and imaging
- Public Health, Environmental and Occupational Health