Efficiency calibration of Ge detectors in wide energy range with natural thorium and 137Cs

Z. B. Alfassi, A. Tsechanski

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

For the efficiency calibration of Ge detectors the 232Th natural radionuclide chain is suggested with a standard source of 137Cs which due to its long half-life requires a change of the calibrated standard only in about 30 to 50 years.

Original languageEnglish
Pages (from-to)309-312
Number of pages4
JournalJournal of Radioanalytical and Nuclear Chemistry
Volume246
Issue number2
DOIs
StatePublished - 25 Nov 2000

ASJC Scopus subject areas

  • Analytical Chemistry
  • Nuclear Energy and Engineering
  • Radiology Nuclear Medicine and imaging
  • Pollution
  • Spectroscopy
  • Public Health, Environmental and Occupational Health
  • Health, Toxicology and Mutagenesis

Fingerprint

Dive into the research topics of 'Efficiency calibration of Ge detectors in wide energy range with natural thorium and 137Cs'. Together they form a unique fingerprint.