Abstract
A high-pressure chamber that can be used inside the sample chamber of a spectrofluorometer is described and some performance characteristics are presented. The chamber body, constructed of 316 stainless steel, is temperature regulated using resistive heating elements and a microprocessor-based proportional integral derivative controller. The chamber holds a standard 1-cm2 cuvette that indexes with an electromagnetic stirrer. Injection of different solutions into the closed and pressurized (6.8 MPa) vessel is accomplished by computer-controlled, low-volume solenoids attached to separate microliter injection ports. Repetitive injections of fluids down to a volume of 7 microliters are possible in the pressurized chamber. Temperature stability of the chamber is +/- 0.2 degrees C at atmospheric or elevated pressure. However, during the initial phase (first 3 min) of pressurization, at a compression rate of 0.62 MPa/min, a 0.23 degrees C/min increase in temperature occurs. The chamber windows depress the relative intensity of the emitted light by approximately 20% for visible light and 40% for near UV; however, total sensitivity of the system is sufficient to accomplish most determinations while maintaining a good signal-to-noise ratio. This system can be used to evaluate the response of several molecular and cellular events during compression and at depth with the use of various fluorometric probes.
| Original language | English |
|---|---|
| Pages (from-to) | 375-382 |
| Number of pages | 8 |
| Journal | Undersea and Hyperbaric Medicine |
| Volume | 20 |
| Issue number | 4 |
| State | Published - 1 Jan 1993 |
| Externally published | Yes |
ASJC Scopus subject areas
- Pulmonary and Respiratory Medicine
- Physiology (medical)
Fingerprint
Dive into the research topics of 'Evaluation of a hyperbaric system to be used in conjunction with a fluorometer.'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver