Abstract
Three types of thermoluminescence dosemeters (TLDs): LiF:Mg,Ti (TLD- 100), CaF2,:Tm (TLD-300), and α- Al2,O3,:C (TLD-500). were investigated for their glow curve response to separate X ray and proton irradiations. The glow curve structure for each individual TLD's exposure to the X ray and proton irradiations was analysed and compared. Distinguishable differences between the glow curve structure characteristic of each type of radiation were observed. The proton TLD-100 glow curve has revealed a complex high-temperature peak structure that was used for the proton/X ray discrimination algorithm. Proton irradiation of TLD-300 resulted in an apparent switch in the relative heights of peaks 3 and 5 as compared to X ray. In TLD-500, proton irradiation produced a more subtle difference in the glow curve with an increase in the ratio between high- and low-temperature peaks. Results demonstrate promising differences in glow curve structure present allowing for discrinfination between X ray and proton radiation field expsoures.
Original language | English |
---|---|
Pages (from-to) | 99-102 |
Number of pages | 4 |
Journal | Radiation Protection Dosimetry |
Volume | 101 |
Issue number | 1-4 |
DOIs | |
State | Published - 1 Jan 2002 |
ASJC Scopus subject areas
- Radiation
- Radiological and Ultrasound Technology
- Radiology Nuclear Medicine and imaging
- Public Health, Environmental and Occupational Health