Abstract
High resolution emission spectra of several TL materials, that are commonly used in dosimetry, were measured using a low cost fibre optic spectrometer containing a 2048 pixel CCD array. The spectra were taken from 177 to 890 nm with a wavelength resolution of 1.3 nm. This allowed for line width discrimination and the resolution of transitions that have not been seen so far. For rare earth doped materials like CaF2:Tm (TLD-300) and CaF2:Dy (TLD-200) the spectral lines were compared to the energy levels measured by Dieke and Crosswhite leading to the identification of most of the transition lines.
Original language | English |
---|---|
Pages (from-to) | 211-216 |
Number of pages | 6 |
Journal | Radiation Protection Dosimetry |
Volume | 100 |
Issue number | 1-4 |
DOIs | |
State | Published - 1 Jan 2002 |
Externally published | Yes |
ASJC Scopus subject areas
- Radiation
- Radiological and Ultrasound Technology
- Radiology Nuclear Medicine and imaging
- Public Health, Environmental and Occupational Health