N2 - A 56-year-old man underwent a single lead Lumax 640 DX implantable cardioverter defibrillator implantation for primary prevention of sudden cardiac death. A DX system consists of a single lead, which provides atrial as well as ventricular electrograms, and enhances atrial arrhythmia detection. Three months after the implantation, high-frequency episodes were detected on the far field and the atrial channels, but not on the bipolar right ventricular channel; these were classified as atrial tachycardia. In the present report, we discussed the unusual pattern of the artifacts that was related to an electromagnetic interference detected by the novel DX system.
AB - A 56-year-old man underwent a single lead Lumax 640 DX implantable cardioverter defibrillator implantation for primary prevention of sudden cardiac death. A DX system consists of a single lead, which provides atrial as well as ventricular electrograms, and enhances atrial arrhythmia detection. Three months after the implantation, high-frequency episodes were detected on the far field and the atrial channels, but not on the bipolar right ventricular channel; these were classified as atrial tachycardia. In the present report, we discussed the unusual pattern of the artifacts that was related to an electromagnetic interference detected by the novel DX system.
KW - Electromagnetic interference (EMI)
KW - Lumax DX ICD
UR - https://www.scopus.com/pages/publications/84955557266
U2 - 10.1016/j.joa.2015.12.004
DO - 10.1016/j.joa.2015.12.004
M3 - Article
AN - SCOPUS:84955557266
SN - 1880-4276
VL - 32
SP - 218
EP - 219
JO - Journal of Arrhythmia
JF - Journal of Arrhythmia
IS - 3
ER -