Improved neutron detection by gamma-ray spectroscopy

Z. B. Alfassi, T. Zlatin, O. Manor, S. Dubinsky, U. German

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

Most high-efficiency interactions for neutron detection are with thermal neutrons, and the detection of the prompt gamma radiation emitted as a result of thermal neutron absorption in specific materials was sometimes used. These materials are named converters, the most widespread being 10B, 113Cd or 73Ge (the last two materials are detector materials themselves for CdZnTe and HPGe detectors). However, for these materials, the main gamma-ray energies are relatively low (in the region 473-600 keV), and there are interferences from other gamma rays present in background. Different combinations of moderator-converter-detector systems were employed to optimise the system efficiency and signal-to-background ratio, using B, Cd, Gd and Cl converters and HPGe, NaI(TI) and BGO detectors. When using a Cl converter, the 1164.7 keV peak seems to be a good choice when a HPGe spectrometer is used. A very practical material containing chlorine is PVC.

Original languageEnglish
Pages (from-to)207-211
Number of pages5
JournalRadiation Protection Dosimetry
Volume110
Issue number1-4
DOIs
StatePublished - 7 Oct 2004

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology
  • Radiation
  • Radiology Nuclear Medicine and imaging
  • Public Health, Environmental and Occupational Health

Fingerprint

Dive into the research topics of 'Improved neutron detection by gamma-ray spectroscopy'. Together they form a unique fingerprint.

Cite this