Inappropriate Shocks and Power Delivery Using Adult Automatic External Defibrillator Pads in a Pediatric Patient

  • Shai Tejman-Yarden
  • , Uriel Katz
  • , Marina Rubinstein
  • , Yehuda Attias
  • , Reem Yahia
  • , David Mishali
  • , Michael Glikson

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

The use of automatic external defibrillators (AEDs) during pulseless resuscitations is considered safe and reliable, and was established as part of the guidelines in out-of-hospital events. Based on extensive studies, the use of the standard AED is now indicated in every age group with a preference of pediatric pad application for small children and babies. If unavailable, adult pads are recommended. We report a case of 2 inappropriate AED shocks that were delivered to a neonate during a pulseless resuscitation after application of adult pads. The 3.6-kg patient received 2 shocks, over 200 J each, for sinus bradycardia that was not detected by the device. Although treated inappropriately with high voltage, no cardiac or skin sequelae were detected, and the patient had normal cardiac and neurological development later on.

Original languageEnglish
Pages (from-to)E1708-E1710
JournalPediatric Emergency Care
Volume37
Issue number12
DOIs
StatePublished - 1 Dec 2021
Externally publishedYes

Keywords

  • Automated external defibrillator
  • Inappropriate shock
  • Under sensing