Abstract
A method for the measurement of a minute activity of a short-lived nuclide embedded in longer-lived higher activity matrix was developed. The method is based on the findings that the ratio of the activities of two γ lines is independent of the dead-time. Another method of mathematical correction was also checked and found to be quite accurate.
Original language | English |
---|---|
Pages (from-to) | 190-192 |
Number of pages | 3 |
Journal | Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |
Volume | 365 |
Issue number | 1 |
DOIs | |
State | Published - 1 Nov 1995 |
ASJC Scopus subject areas
- Nuclear and High Energy Physics
- Instrumentation