Measurements of laser generated soft X-ray emission from irradiated gold foils

J. S. Davis, Y. Frank, E. Raicher, M. Fraenkel, P. A. Keiter, S. R. Klein, R. P. Drake, D. Shvarts

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

Soft x-ray emission from laser irradiated gold foils was measured at the Omega-60 laser system using the Dante photodiode array. The foils were heated with 2 kJ, 6 ns laser pulses and foil thicknesses were varied between 0.5, 1.0, and 2.0 μm. Initial Dante analysis indicates peak emission temperatures of roughly 100 eV and 80 eV for the 0.5 μm and 1.0 μm thick foils, respectively, with little measurable emission from the 2.0 μm foils.

Original languageEnglish
Article number11D609
JournalReview of Scientific Instruments
Volume87
Issue number11
DOIs
StatePublished - 1 Nov 2016