TY - JOUR
T1 - Mobile ultra-clean unidirectional airflow screen reduces air contamination in a simulated setting for intra-vitreal injection
AU - Lapid-Gortzak, Ruth
AU - Traversari, Roberto
AU - van der Linden, Jan Willem
AU - Lesnik Oberstein, Sarit Y.
AU - Lapid, Oren
AU - Schlingemann, Reinier O.
N1 - Publisher Copyright:
© 2016, The Author(s).
PY - 2017/2/1
Y1 - 2017/2/1
N2 - The aim of this study is to determine whether the use of a mobile ultra-clean laminar airflow screen reduces the air-borne particle counts in the setting of a simulated procedure of an intra-vitreal injection. A mobile ultra-clean unidirectional airflow (UDF) screen was tested in a simulated procedure for intra-vitreal injections in a treatment room without mechanical ventilation. One UDF was passed over the instrument tray and the surgical area. The concentration of particles was measured in the background, over the instrument table, and next to the ocular area. The degree of protection was calculated at the instrument table and at the surgical site. Use of the UDF mobile screen reduced the mean particle concentration (particles > 0.3 microns) on the instrument table by a factor of at least 100.000 (p < 0.05), and over the patient’s eye by at least a factor of 436 (p < 0.05), which in clinical practice translates into significantly reduced air contamination. Mobile UDF screen reduces the mean particle concentration substantially. The mobile UDF screen may therefore allow for a safer procedural environment for ambulatory care procedures such as intra-vitreal injections in treatment rooms.
AB - The aim of this study is to determine whether the use of a mobile ultra-clean laminar airflow screen reduces the air-borne particle counts in the setting of a simulated procedure of an intra-vitreal injection. A mobile ultra-clean unidirectional airflow (UDF) screen was tested in a simulated procedure for intra-vitreal injections in a treatment room without mechanical ventilation. One UDF was passed over the instrument tray and the surgical area. The concentration of particles was measured in the background, over the instrument table, and next to the ocular area. The degree of protection was calculated at the instrument table and at the surgical site. Use of the UDF mobile screen reduced the mean particle concentration (particles > 0.3 microns) on the instrument table by a factor of at least 100.000 (p < 0.05), and over the patient’s eye by at least a factor of 436 (p < 0.05), which in clinical practice translates into significantly reduced air contamination. Mobile UDF screen reduces the mean particle concentration substantially. The mobile UDF screen may therefore allow for a safer procedural environment for ambulatory care procedures such as intra-vitreal injections in treatment rooms.
KW - Contamination
KW - Endophthalmitis prevention
KW - HEPA filters
KW - Intra-vitreal injections
KW - Unidirectional ultra-clean laminar airflow
UR - http://www.scopus.com/inward/record.url?scp=84965014210&partnerID=8YFLogxK
U2 - 10.1007/s10792-016-0236-1
DO - 10.1007/s10792-016-0236-1
M3 - Article
C2 - 27138593
AN - SCOPUS:84965014210
SN - 0165-5701
VL - 37
SP - 131
EP - 137
JO - International Ophthalmology
JF - International Ophthalmology
IS - 1
ER -