Reassessment of doses in TLD-100 after long storage times

A. Abraham, M. Weinstein, U. German, Z. B. Alfassi, Z. Machlev

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

Reassessment of doses measured by TLD-100 dosimeters can be performed by extended time readout (for 30 s) or by using the photo-transferred thermoluminscence (PTTL) method. If the reassessment is performed after a long storage time, or after storage in high background environments, the background buildup may interfere with the reassessment results, inducing significant errors. An experimental procedure was applied, which enables reassessment of doses down to about 10 mSv, even for very high background contributions. No measurable fading effect could be observed.

Original languageEnglish
Pages (from-to)802-804
Number of pages3
JournalRadiation Measurements
Volume43
Issue number2-6
DOIs
StatePublished - 1 Feb 2008

Keywords

  • PTTL
  • Reassessment
  • Residual dose
  • Storage