The effect of sample/planchet geometry and temperature resolution on the reproducibility of glow curve shapes and precision of dose measurement in LiF-TLD-100 thermoluminescent dosimetry

D. Sibony, Y. Horowitz, L. Oster

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

The effect of accurate positioning of TLD-100 samples in the center of the reader planchet on precision and kinetic parameters was investigated. Significant improvement in precision is obtained by careful positioning of the sample in the center of the planchet, by as much as 40%, 20% and 30% using different methods of glow curve analysis to estimate the intensity of the TL signal.

Original languageEnglish
Pages (from-to)205-207
Number of pages3
JournalRadiation Measurements
Volume71
DOIs
StatePublished - 1 Jan 2014

Keywords

  • Planchet geometry
  • Precision of dose measurements
  • TL glow curve shapes
  • TLD-100