Thermoluminescent LiF:Cu,Mg,P for gamma ray dosimetry in mixed fast neutron-gamma radiation fields

Y. S. Horowitz, B. B. Shachar

Research output: Contribution to journalArticlepeer-review

25 Scopus citations

Abstract

In thermoluminescent dosimetry (TLD), the relative TL response, η(nγ), of fast neutrons to gamma rays usually lies between 0.05 to 0.3. Recently, several groups have reported on the development of a new TL material, LiF:Cu,Mg,P, with a gamma ray TL sensitivity ~20 times greater than LiF:Mg,Ti(TLD-100), Harshaw-Filtrol). We have measured the relative TL response of LiF:Cu,Mg,P to 252Cf fast neutrons (Ēn) = 2.5 MeV). Neutron dosimetry was carried out using the polycarbonate technique, and, for comparison purposes, the neutron irradiations were also carried out simultaneously on pairs of LiF:Mg,Ti(TLD-700, H/F) and CaF2:Tm (TLD-300, H/F) dosemeters. The following results were obtained: (1) n(LiF:Cu,Mg,P) = 0.011 ± 0.01; (2) n(TLD-700)/n(LiF:Cu,Mg,P) ≃5:1; and (3) n(TLD-300)/n(LiF:Cu,Mg,P) ≃ 4:1. Estimation of the TL signal was carried out using computerised glow curve deconvolution (CGCD). These results establish LiF:Cu,Mg,P as the 'tissue-equivalent' TL material least sensitive to fast neutrons discovered to date and suggest that LiF:Cu,Mg,P can serve as an excellent discriminator for gamma ray dose measurement in fast neutron - gamma ray radiation fields.

Original languageEnglish
Pages (from-to)401-404
Number of pages4
JournalRadiation Protection Dosimetry
Volume23
Issue number1-4
DOIs
StatePublished - 1 Jan 1988

ASJC Scopus subject areas

  • Radiation
  • Radiological and Ultrasound Technology
  • Radiology Nuclear Medicine and imaging
  • Public Health, Environmental and Occupational Health

Fingerprint

Dive into the research topics of 'Thermoluminescent LiF:Cu,Mg,P for gamma ray dosimetry in mixed fast neutron-gamma radiation fields'. Together they form a unique fingerprint.

Cite this