Abstract
A fast neutron detector (En>3 MeV) was constructed by covering a Ge(Li) diode with teflon [(CF2)n] and detecting the 6. 131 MeV γ line of 16N obtained from the 19F(n, α) reaction. Possible uses of this detector are given.
Original language | English |
---|---|
Pages (from-to) | 195-197 |
Number of pages | 3 |
Journal | Nuclear Instruments and Methods |
Volume | 148 |
Issue number | 1 |
DOIs | |
State | Published - 1 Jan 1978 |