TY - JOUR
T1 - Validation study to determine the accuracy of central blood pressure measurement using the SphygmoCor XCEL cuff device in patients with severe aortic stenosis undergoing transcatheter aortic valve replacement
AU - De la Torre Hernández, Jose M.
AU - Veiga Fernandez, Gabriela
AU - Brown, Jonathan
AU - Sainz Laso, Fermin
AU - Lee, Dae Hyun
AU - Fradejas, Victor
AU - Garcia Camarero, Tamara
AU - Elmariah, Sammy
AU - Inglessis, Ignacio
AU - Zueco, Javier
AU - Vazquez de Prada, Jose A.
AU - Ben-Assa, Eyal
AU - Edelman, Elazer R.
N1 - Publisher Copyright:
© © 2021 The Authors. The Journal of Clinical Hypertension published by Wiley Periodicals LLC.
PY - 2021/6/1
Y1 - 2021/6/1
N2 - Central aortic blood pressure could be helpful in the evaluation of patients with aortic stenosis (AS). The SphygmoCor XCEL device estimates central blood pressure (BP) measurement with its easy-to-use, operator-independent procedure. However, this device has not been properly validated against invasive measurement in patients with severe AS. We evaluated the relationship between cuff-brachial BP, transfer function-estimated and invasively measured central aortic pressure in patients with severe AS before and after transcatheter aortic valve replacement (TAVR). Agreement between techniques was analyzed and, according to the ARTERY Society recommendations, the minimum acceptable error was a mean difference ± SD ≤5 ± ≤8 mm Hg. A total of 94 patients with AS undergoing TAVR had simultaneous non-invasive and invasive measurements of central BP before and after the procedure. Before TAVR central systolic BP was in average slightly underestimated, though with wide variability, when using the default calibration of brachial-cuff SBP (mean difference ± SD, −3 ± 15 mm Hg), and after TAVR the degree of underestimation increased (mean difference ± SD, −9 ± 13 mm Hg). The agreement tended to improve for those patients with low aortic gradient stenosis compared to those with high gradient at baseline (mean difference ± SD, −2 ± 11 mm Hg vs. −4 ± 17, respectively, p =.3). The cuff-brachial systolic BP yielded numerically lower degree of agreement and weaker correlation with invasive measurements than SphygmoCor XCEL. In patients with severe AS the SphygmoCor XCEL cuff device, despite showing strong correlation, does not meet the ARTERY Society accuracy criteria for non-invasive measurement of central SBP.
AB - Central aortic blood pressure could be helpful in the evaluation of patients with aortic stenosis (AS). The SphygmoCor XCEL device estimates central blood pressure (BP) measurement with its easy-to-use, operator-independent procedure. However, this device has not been properly validated against invasive measurement in patients with severe AS. We evaluated the relationship between cuff-brachial BP, transfer function-estimated and invasively measured central aortic pressure in patients with severe AS before and after transcatheter aortic valve replacement (TAVR). Agreement between techniques was analyzed and, according to the ARTERY Society recommendations, the minimum acceptable error was a mean difference ± SD ≤5 ± ≤8 mm Hg. A total of 94 patients with AS undergoing TAVR had simultaneous non-invasive and invasive measurements of central BP before and after the procedure. Before TAVR central systolic BP was in average slightly underestimated, though with wide variability, when using the default calibration of brachial-cuff SBP (mean difference ± SD, −3 ± 15 mm Hg), and after TAVR the degree of underestimation increased (mean difference ± SD, −9 ± 13 mm Hg). The agreement tended to improve for those patients with low aortic gradient stenosis compared to those with high gradient at baseline (mean difference ± SD, −2 ± 11 mm Hg vs. −4 ± 17, respectively, p =.3). The cuff-brachial systolic BP yielded numerically lower degree of agreement and weaker correlation with invasive measurements than SphygmoCor XCEL. In patients with severe AS the SphygmoCor XCEL cuff device, despite showing strong correlation, does not meet the ARTERY Society accuracy criteria for non-invasive measurement of central SBP.
KW - aortic stenosis
KW - central aortic blood pressure
KW - hypertension
KW - transcatheter aortic valve replacement
UR - http://www.scopus.com/inward/record.url?scp=85105043249&partnerID=8YFLogxK
U2 - 10.1111/jch.14245
DO - 10.1111/jch.14245
M3 - Article
C2 - 33942487
AN - SCOPUS:85105043249
SN - 1524-6175
VL - 23
SP - 1165
EP - 1175
JO - Journal of Clinical Hypertension
JF - Journal of Clinical Hypertension
IS - 6
ER -